用户名: 密码: 验证码:
New material for orthopedic implants: Electrochemical study of nickel free P558 stainless steel in minimum essential medium
详细信息    查看全文
文摘
Nickel, a component of stainless steels (SS) applied in orthopedic implants may cause allergic processes in human tissues. P558 nickel free SS was studied to verify its viability as a substitute for stainless steel containing nickel. Its performance is compared to ISO 5832-9 and F138 most used nowadays grades in implants fabrications, in minimum essential medium, MEM, at 37 °C. Potentiodynamic polarization curves, electrochemical impedance spectroscopy (EIS), scanning electron microscopy (SEM) and “in vitro” cytotoxicity were used as techniques. From the electrochemical point of view P558 SS is comparable to ISO 5832-9 SS in MEM. It remains passivated until the transpassivation potential, above which generalized corrosion occurs. F138 presents pitting corrosion at 370 mV/SCE. The cytotoxicity results showed that P558, ISO 5832-9 and F138 do not present cytotoxic character. Therefore, these results suggest that P558 SS can be applied in orthopedic implants.