A systematic comparison of four bioaerosol generators: Affect on culturability and cell membrane integrity when aerosolizing Escherichia coli bacteria
详细信息    查看全文
文摘
Bioaerosol research requires stable and reliable aerosol generators that can produce high particle concentrations with minimal damage to microorganisms. This study compared Collison nebulizer, Liquid Sparging Aerosolizer (LSA), C-Flow nebulizer, and a newly designed Single-Pass Aerosolizer with respect to their physical performance, and ability to preserve the culturability and structural integrity of bacteria. Escherichia coli bacteria were aerosolized at different air pressures, collected by a BioSampler and their Cell Membrane Damage Index (ID), expressed as the fraction of 16S rRNA gene copies in the supernatant liquid versus the amount of 16S rRNA gene copies in the total sample (cell pellet plus supernatant), was determined. The ID of E. coli aerosolized by the Collison and C-Flow nebulizers at 40 psi compared to aerosolization at 5 and 15 psi was significantly higher (p<0.05). However, the ID of E. coli aerosolized with the LSA and Single-Pass Aerosolizer did not seem to significantly depend on aerosolization pressure. The ID of E. coli collected with a BioSampler was found to positively and significantly correlate (p=0.043) with the presence of airborne bacterial fragments (aerodynamic size range 0.37-0.523 碌m) as measured by the Aerodynamic Particle Sizer. Increased loss of culturability was observed for bacteria aerosolized by the Collison nebulizer and Single-Pass Aerosolizer with increasing aerosolization pressure (p<0.05), while no significant change in culturability was found for the other two generators as a function of aerosolization pressure. At particle output concentration of ~100 particles/cm3, the Single-Pass Aerosolizer preserved the culturability of bacteria significantly better than the other three generators (p<0.05). It also exhibited a significantly lower ID (p<0.001) and less culturability reduction (p=0.03) compared to the Collison nebulizer at particle output concentrations of ~1000 particles/cm3. It is hoped that this study will help bioaerosol researchers select a bioaerosol generator and method best suiting their studies.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700