ABG activity was measured in 3.2 mm punches from dry blood spots (DBS). Each punch was incubated for 21 h with the substrate D-Glucosyl-β1-1′-N-dodecanoyl-D-erythro-sphingosine [C12-glucocerebroside (C36H69NO8)] and internal standard N-myristoyl-D-erythro-sphingosine [C14-ceramide (C32H63NO3)]. The product and internal standard were quantified using MS/MS.
ABG activities in anonymized newborn screening samples from NY State were (mean) 22.0 μmol/h/L ± (SD) 13.8 μmol/h/L (n = 2088, median 19.9 μmol/h/L, 95 % CI 22.59–21.41 μmol/h/L). The enzymatic activity in DBS from 10 treatment naïve adult Gaucher patients was less than 4.2 μmol/h/L. ABG activity was stable for 3 months at room temperature a 20 % activity reduction was observed. Inter- and intra-run imprecisions were 8 % and 13.7 % , respectively. The limit of detection was 0.75 μmol/h/L and limit of quantification was 1.25 μmol/h/L.
The measurement of ABG activities in DBS using MS/MS is suitable for high-throughput analysis of at-risk individuals and potentially for newborn screening for GD.