Pacemaker Stimulus Amplitude Alteration without Loss of Capture: An Unusual ECG Finding in Cardiac Tamponade from Pacemaker Lead Perforation
详细信息    查看全文
文摘
A variation in pacemaker stimulus amplitude can represent pacemaker system dysfunction from generator malfunction, lead insulation defect, lead fracture, or artefact of digital signal processing of the electrocardiography recorder. Pacemaker lead perforation into the pericardial space typically results in loss of capture which was not demonstrated in our patient. In summary, we report an unusual ECG finding of pacemaker stimulus amplitude alteration without loss of capture in the setting of cardiac tamponade from pacemaker lead perforation.