用户名: 密码: 验证码:
Evaluation of the use of a Laser Tracker and an Indexed Metrology Platform as Gauge Equipment in Articulated Arm Coordinate Measuring Machine Verification Procedures
详细信息    查看全文
文摘
The development of new gauges and alternative calibration and verification techniques applied to articulated arm coordinate measuring machine (AACMM) has been approached in different ways. In this work the analysis of the use of a laser tracker (LT) as reference instrument in calibration and verification procedures for AACMM together with an indexed metrology platform (IMP) replacing the conventional one-dimensional gauges is presented. This technique avoids the need of materializing the length required in a conventional gauge, increasing the flexibility for defining test positions and broadening the definition of reference lengths depending on equipment calibration or verification requirements. The use of the IMP in this verification procedure enables a reduction of testing time and space needed in the data capturing process. The results obtained showed the suitability of the proposed procedure using the LT as reference instrument.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700