120 patients with a suspicion of lung cancer were enrolled prospectively. PET images were reconstructed with the non-enhanced and venous phase contrast CT obtained after injection of iopromide 300 mg/ml or 370 mg/ml using either a fixed-dose or a body surface area adapted injection protocol. Maximum and mean standardized uptake values (SUVmax and SUVmean) and contrast enhancement (HU) were determined in the subclavian vein, ascending aorta, abdominal aorta, inferior vena cava, portal vein, liver and kidney and in the suspicious lung lesion. PET data were evaluated visually for the presence of malignancy and image quality.
At none of the sites a significant difference in the extent of the contrast enhancement between the four different protocols was found. However, the variability of the contrast enhancement at several anatomical sites was significantly greater in the fixed dose groups than in the BSA groups for both contrast medium concentrations. At none of the sites a significant difference was found in the extent of the SUVmax and SUVmean increase as a result of the use of the venous phase contrast enhanced CT for attenuation. Visual clinical evaluation of lesions showed no differences between contrast and non-contrast PET/CT (P = 0.32).
Contrast enhanced CT for attenuation correction in combined PET/CT in lung cancer affects neither the clinical assessment nor image quality of the PET-images. A body surface adapted contrast medium protocol reduces the interpatient variability in contrast enhancement.