用户名: 密码: 验证码:
Transcatheter Closure of Patent Ductus Arteriosus Using the Amplatzer Duct Occluder in Infants Under 1 Year of Age
详细信息    查看全文
文摘

Introduction and objectives

Percutaneous closure of patent ductus arteriosus (PDA) is a well-established technique. We evaluated the usefulness of the Amplatzer duct occluder for the percutaneous closure of patent ductus arteriosus in 29 children under 1 year of age.

Methods

The patients’ mean age was 8.9 [2.8] months and their mean weight was 6.4 [1.5] kg. In addition, 24.1 % of patients were aged 6 months or less and 17.2 % weighed 5 kg or less. All completed follow-up (0.5-36 months).

Results

The minimum PDA diameter was 3.16[1.24] mm. The device was implanted successfully in 26 patients (89.6 % ). Failures were due to either the device migrating to the descending aorta, persistent moderate leakage, or to difficulty in the advancement of the device. No deaths were associated with the procedure. Three major complications occurred in two patients (10.3 % ). Aortography showed initial total occlusion in 65.5 % of patients. At 3-month follow-up, total occlusion was observed in 96.1 % of patients. Ultimately, closure was achieved successfully in 25 of the 29 patients (86.2 % ). During follow-up, 4 patients exhibited mild left pulmonary artery stenosis and 1 exhibited mild stenosis of the descending aorta.

Conclusions

In children under 1 year of age, percutaneous closure of patent ductus arteriosus using an Amplatzer occluder was a safe and effective procedure. It is possible that improvements in the design of the occluder could decrease the complication rate.