There is an increasing interest in transcatheter mitral valve (MV) treatment. The NeoChord DS 1000 system enables off-pump beating heart transapical MV repair with implantation of artificial neo-chordae.
Patients with severe mitral regurgitation (MR) due to isolated posterior prolapse were included in this TACT (Transapical Artificial Chordae Tendinae) trial. All patients were scheduled for off-pump transapical implantation of neo-chordae.
Thirty patients at 7 centers were enrolled. Major adverse events included 1 death due to post-cardiotomy syndrome and concomitant sepsis and 1 minor stroke with the patient fully recovered at the 30-day follow-up visit. Additional patients experienced procedural major adverse events related to a reoperation or conversion to standard of care. Acute procedural success (placement of at least 1 neo-chord and reduction of MR from 3+ or 4+ to聽鈮?+) was achieved in聽26 patients (86.7%). In 4 patients neo-chordae were not placed for technical and/or patient-specific reasons. These patients underwent intraoperative (3 patients) or post-operative (1 patient) standard MV repair. At 30 days, 17 patients maintained an MR grade聽鈮?+. Four patients who developed recurrent MR were successfully treated with open MV repair during 30-day follow-up. Results improved with experience: durable reduction in MR to聽鈮?+ at 30 days was achieved in 5 (33.3%) of the first 15 patients and 12 (85.7%) of the last 14 patients.
Off-pump transapical implantation of artificial chordae to correct MR is technically safe and feasible; however, it yields further potential for improvement of efficacy and durability. (Safety and Performance Study of the NeoChord Device [TACT]; )