用户名: 密码: 验证码:
Application of the dual-frequency ultrasonometer for osteoporosis detection
详细信息    查看全文
文摘
The paper presents results of a clinical validation study of Bone UltraSonic Scanner (BUSS), a novel dual-frequency axial transmission ultrasonometer, developed by Artann Laboratories. Assessment of bone conditions is based on evaluating relative changes of the axial profiles of ultrasonic characteristics in long bones and utilizes bulk and guided acoustic waves. The objective of this study was to determine the ability of BUSS to discriminate osteoporosis development stages. A total of 93 menopausal and post-menopausal women divided into five groups from normal to advanced osteoporosis according to their DXA hip t-score were enrolled in the study. The 2D waveform profiles at low (0.1 MHz) and high (1 MHz) frequencies were obtained by scanning 15 cm along the proximal tibia.

A multi-parametric linear classifier based on a set of the parameters derived from 2D acoustic waveform profiles has been developed. The efficiency of this classifier in differentiating osteoporosis from a normal sample was assessed using a receiver operating characteristic (ROC) curve analysis. Based on the ROC analysis, BUSS demonstrated 76 % sensitivity and 70 % specificity to DXA-identified osteoporosis. The area under the ROC curve, which is a measure of how well a parameter can distinguish between the two diagnostic groups (diseased/normal) was 79.3 % . The study confirmed BUSS’s capability to discriminate between stages of bone atrophy and in particular to distinguish early changes induced by osteoporosis.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700