Comparison of advanced iterative reconstruction methods for SPECT/CT
详细信息    查看全文
文摘

Aim

Corrective image reconstruction methods which produce reconstructed images with improved spatial resolution and decreased noise level became recently commercially available. In this work, we tested the performance of three new software packages with reconstruction schemes recommended by the manufacturers using physical phantoms simulating realistic clinical settings.

Methods

A specially designed resolution phantom containing three 99mTc lines sources and the NEMA NU-2 image quality phantom were acquired on three different SPECT/CT systems (General Electrics Infinia, Philips BrightView and Siemens Symbia T6). Measurement of both phantoms was done with the trunk filled with a 99mTc-water solution. The projection data were reconstructed using the GE's Evolution for Bone?, Philips Astonish? and Siemens Flash3D? software. The reconstruction parameters employed (number of iterations and subsets, the choice of post-filtering) followed theses recommendations of each vendor. These results were compared with reference reconstructions using the ordered subset expectation maximization (OSEM) reconstruction scheme.

Results

The best results (smallest value for resolution, highest percent contrast values) for all three packages were found for the scatter corrected data without applying any post-filtering. The advanced reconstruction methods improve the full width at half maximum (FWHM) of the line sources from 11.4 to 9.5 mm (GE), from 9.1 to 6.4 mm (Philips), and from 12.1 to 8.9 mm (Siemens) if no additional post filter was applied. The total image quality control index measured for a concentration ratio of 8:1 improves for GE from 147 to 189, from 179. to 325 for Philips and from 217 to 320 for Siemens using the reference method for comparison. The same trends can be observed for the 4:1 concentration ratio. The use of a post-filter reduces the background variability approximately by a factor of two, but deteriorates significantly the spatial resolution.

Conclusions

Using advanced reconstruction algorithms the largest improvement in image resolution and contrast is found for the scatter corrected slices without applying post-filtering. The user has to choose whether noise reduction by post-filtering or improved image resolution fits better a particular imaging procedure.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700