用户名: 密码: 验证码:
Integrated SFE/SFC/MS System for the Analysis of Pesticides in Animal Tissues
详细信息    查看全文
  • 作者:Kent J. Voorhees ; Ahmad A. Gharaibeh ; and B. Murugaverl
  • 刊名:Journal of Agricultural and Food Chemistry
  • 出版年:1998
  • 出版时间:June 1998
  • 年:1998
  • 卷:46
  • 期:6
  • 页码:2353 - 2359
  • 全文大小:123K
  • 年卷期:v.46,no.6(June 1998)
  • ISSN:1520-5118
文摘
The development of an integrated system for an on-line supercriticalfluid extraction/cleanup/supercritical fluid chromatography/mass spectrometry analysis(SFE/SFC/MS) of thermally labilecarbamate pesticides from beef and chicken meat samples is discussed.A key to the system is themixed stationary phase (7% diol in C18) used in thecleanup column to separate target analytesfrom the extracted lipid matrix interferences. The detectionlimits of the system at 2:1 signal tonoise ratio were 200 and 175 ppb for bendiocarb and carbaryl,respectively. The reproducibility ofthe present system was not totally favorable due to trapping problems,resulting from frequentclogging of either the cryogenic retention gap between the extractioncell and the SFC or the actualrestrictor in the retention gap. Operating in an integrated mode,53% of the experiments werecompleted without complications connected to extracting and detectingthe analyte.Keywords: Carbamates; pesticides; SFE; SFC; integrated system; on-linecleanup; animal tissues

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700