用户名: 密码: 验证码:
Validation and Application of Cavity-Enhanced, Near-Infrared Tunable Diode Laser Absorption Spectrometry for Measurements of Methane Carbon Isotopes at Ambient Concentrations
详细信息    查看全文
文摘
Methane is an effective greenhouse gas but has a short residence time in the atmosphere, and therefore, reductions in emissions can alleviate its greenhouse gas warming effect within a decadal time frame. Continuous and high temporal resolution measurements of methane concentrations and carbon isotopic ratios (未13CH4) can inform on mechanisms of formation, provide constraints on emissions sources, and guide future mitigation efforts. We describe the development, validation, and deployment of a cavity-enhanced, near-infrared tunable diode laser absorption spectrometry system capable of quantifying 未13CH4 at ambient methane concentrations. Laboratory validation and testing show that the instrument is capable of operating over a wide dynamic range of methane concentration and provides a measurement precision for 未13CH4 of better than 卤0.5鈥?(1蟽) over 1000 s of data averaging at ambient methane concentrations. The analyzer is accurate to better than 卤0.5鈥? as demonstrated by measurements of characterized methane/air samples with minimal dependence (<1鈥? of measured carbon isotope ratio on methane concentration. Deployment of the instrument at a marsh over multiple days demonstrated how methane fluxes varied by an order of magnitude over 2 day deployment periods, and showed a 17 鈥?variability in 未13CH4 of the emitted methane during the growing season.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700