用户名: 密码: 验证码:
Simultaneous Identification and Antimicrobial Susceptibility Testing of Multiple Uropathogens on a Microfluidic Chip with Paper-Supported Cell Culture Arrays
详细信息    查看全文
文摘
A microfluidic chip was developed for one-step identification and antimicrobial susceptibility testing (AST) of multiple uropathogens. The polydimethylsiloxane (PDMS) microchip used had features of cell culture chamber arrays connected through a sample introduction channel. At the bottom of each chamber, a paper substrate preloaded with chromogenic media and antimicrobial agents was embedded. By integrating a hydrophobic membrane valve on the microchip, the urine sample can be equally distributed into and confined in individual chambers. The identification and AST assays on multiple uropathogens were performed by combining the spatial resolution of the cell culture arrays and the color resolution from the chromogenic reaction. The composite microbial testing assay was based on dynamic changes in color in a serial of chambers. The bacterial antimicrobial susceptibility was determined by the lowest concentration of an antimicrobial agent that is capable of inhibiting the chromogenic reaction. Using three common uropathogenic bacteria as test models, the developed microfluidic approach was demonstrated to be able to complete the multiple colorimetric assays in 15 h. The accuracy of the microchip method, in comparison with that of the conventional approach, showed a coincidence of 94.1%. Our data suggest this microfluidic approach will be a promising tool for simple and fast uropathogen testing in resource-limited settings.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700