Interface for Online Coupling of Surface Plasmon Resonance to Direct Analysis in Real Time Mass Spectrometry
详细信息    查看全文
文摘
The online coupling of surface plasmon resonance (SPR) with mass spectrometry (MS) has been highly desired for the complementary information provided by each of the two techniques. In this work, a novel interface for direct and online coupling of SPR to direct analysis in real time (DART) MS was developed. A spray tip connected with the outlet of the SPR flow solution was conducted as the sampling part of the DART-MS, with which the online coupling interface of SPR-MS was realized. Four model samples, acetaminophen, metronidazole, quinine, and hippuric acid, dissolved in three kinds of common buffers were used in the SPR-DART-MS experiments for performance evaluation of the interface and the optimization of DART conditions. The results showed consistent signal changes and high tolerance of nonvolatile salts of this SPR-MS system, demonstrating the feasibility of the interface for online coupling of SPR with MS and the potential application in the characterization of interaction under physiological conditions.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700