用户名: 密码: 验证码:
LED-Based UV Absorption Detector with Low Detection Limits for Capillary Liquid Chromatography
详细信息    查看全文
文摘
A 260 nm deep UV LED-based absorption detector with low detection limits was developed and integrated with a small nanoflow pumping system. The detector is small in size (5.2 脳 3.0 cm) and weighs only 85 g (without electronics). This detector was specifically designed and optimized for on-column detection to minimize extra-column band broadening. No optical reference was included due to the low drift in the signal. Two ball lenses, one of which was integrated with the LED, were used to increase light throughput through the capillary column. Stray light was minimized by the use of a band-pass filter and an adjustable slit. Signals down to the parts per billion level (nanomolar) were easily detected with a short-term noise level of 4.4 渭AU, confirming a low limit of detection and low noise. The detection limit for adenosine-5鈥?monophosphate was 230 times lower than any previously reported values. Good linearities (3 orders of magnitude) were obtained using sodium anthraquinone-2-sulfonate, adenosine-5鈥?monophosphate, dl-tryptophan, and phenol. The LC system was demonstrated by performing isocratic separation of phenolic compounds using a monolithic capillary column (16.5 cm 脳 150 渭m i.d.) synthesized from poly(ethylene glycol) diacrylate.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700