文摘
Recently, Monte Carlo methods have been used to optimize the design and modeling of radiation detectors. However, most Monte Carlo codes have a fixed and simple optical physics, and the effect of the signal readout devices is not considered because of the limitations of the geometry function. Therefore, the disadvantages of the codes prevent the modeling of the scintillator detector. The modeling of a comprehensive and extensive detector system has been reported to be feasible when the optical physics model of the GEomerty ANd Tracking 4 (GEANT 4) simulation code is used. In this study, we performed a Gd2O3:Eu scintillator modelling by using the GEANT4 simulation code and compared the results with the measurement data. To obtain the measurement data for the scintillator, we synthesized the Gd2O3:Eu scintillator by using solution combustion method and we evaluated the characteristics of the scintillator by using X-ray diffraction and photoluminescence. We imported the measured data into the GEANT4 code because GEANT4 cannot simulate a fluorescence phenomenon. The imported data were used as an energy distribution for optical photon generation based on the energy deposited in the scintillator. As a result of the simulation, a strong emission peak consistent with the measured data was observed at 611 nm, and the overall trends of the spectrum agreed with the measured data. This result is significant because the characteristics of the scintillator are equally implemented in the simulation, indicating a valuable improvement in the modeling of scintillator-based radiation detectors.