文摘
Background Following the recent development of a new approach to quantitative analysis of IgG concentrations in bovine serum using transmission infrared spectroscopy, the potential to measure IgG levels using technology and a device better designed for field use was investigated. A method using attenuated total reflectance infrared (ATR) spectroscopy in combination with partial least squares (PLS) regression was developed to measure bovine serum IgG concentrations. ATR spectroscopy has a distinct ease-of-use advantage that may open the door to routine point-of-care testing. Serum samples were collected from calves and adult cows, tested by a reference RID method, and ATR spectra acquired. The spectra were linked to the RID-IgG concentrations and then randomly split into two sets: calibration and prediction. The calibration set was used to build a calibration model, while the prediction set was used to assess the predictive performance and accuracy of the final model. The procedure was repeated for various spectral data preprocessing approaches.