Determination of V, Ti, Co and Ni in Vanadium and Titano Magnetite by Inductively Coupled Plasma-Atomic Emission Spectrometry
摘要
Vanadium and titano magnetite was melted by sodium peroxide, then extracted by HCl. The sample solution was diluted before V2O5, TiO2, Co and Ni were simultaneously directly determined by Inductively Coupled Plasma-Atomic Emission Spectrometry (ICP-AES). This method exhibited simplicity, rapidity, good reproducibility and wide linearity making it suitable for multi-elemental determination.