用户名: 密码: 验证码:
Test Quality Risks and Countermeasure of Mineralized Geochemical Samples
详细信息   
摘要
Because of the content of certain element in mineralized sample is high,which causes interference to the measurement of other elements,the quality parameters not out-of-tolerance,there is omissions and false positive risk.Aiming at lead test of 10 mineralized samples of 1:50000 regional geological exploration in Hulunbuir Bajianfang of Inner Mongolia Autonomous Region,performed method comparison experiment.High calcium samples interfered the determination of lead by atomic absorption spectrometry,test results are much higher than the true value.Suggest analyzing geochemical samples in laboratory,should use different analysis methods on internal analysis and abnormal spot check,to avoid false positive and omissions risk.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700