用户名: 密码: 验证码:
Report of the Proficiency Testing in the Pneumatic Pressure Region up to 5?MPa
详细信息    查看全文
  • 作者:Jasveer Singh (1)
    Nita Dilawar Sharma (1)
    Ashok Kumar (1)
    A. K. Bandyopadhyay (1)
  • 关键词:Proficiency testing ; Pneumatic pressure ; Quality control
  • 刊名:Mapan
  • 出版年:2014
  • 出版时间:September 2014
  • 年:2014
  • 卷:29
  • 期:3
  • 页码:213-222
  • 全文大小:976 KB
  • 参考文献:1. ISO/IEC, General requirements for the competence of testing and calibration laboratories, ISO/IEC 17025, (2005).
    2. https://www.ilac.org/ilacarrangement.html. Accessed June 2013
    3. https://www.aplac.org/aplac_mra.html.. Accessed June 2013
    4. S. Yadav and A. K. Bandyopadhyay, Interlaboratory comparison in the pressure range 7-0?MPa using digital pressure calibrator, MAPAN-Journal of Metrology Society of India, 20 (1) (2005) 297-10.
    5. S. Yadav and A. K. Bandyopadhyay, Evaluation of laboratory performance through Interlaboratory comparison, MAPAN-Journal of Metrology society of India, 24 (2) (2009) 125-38.
    6. S. Yadav, V. K. Gupta, and A. K. Bandyopadhyay, Standardization of pressure measurement using pressure balance as transfer standard, MAPAN-Journal of Metrology society of India, 26 (2) (2011), 133-51.
    7. ISO/IEC, Conformity assessment—general requirements for proficiency testing, ISO/IEC 17043, (2010).
    8. NABL, Guidelines for proficiency testing program for testing and calibration laboratories, NABL Doc. 162(4) (2008).
    9. NABL, Directory of accredited calibration laboratories, NABL Doc. 500, (2010).
    10. N. Dilawar Sharma, A. Vijayakumar, J. Singh, A. Kumar, D. R. Sharma, and A. K. Bandyopadhyay, Establishing a continuous chain of traceability for pressure measurements up to 40?MPa, NCSLI Measure J. Meas. Sci., 8 (1) (2013) 56-5.
    11. R. Gregory Driver, D. B. Olson, N Dilawar, and A. K. Bandyopadhyay, Final report on key comparison APMP.SIM.M.P-K1c: bilateral comparison between NIST (USA) and NPLI (India) in the pneumatic pressure region 0.4?MPa to 4.0?MPa, Metrologia Tech. Suppl., 44 (2007) 7002. CrossRef
    12. A. K Bandyopadhyay et al., Results of the APMP pressure key comparison APMP.M.P-K1c in gas media and gauge mode from 0.4 to 4?MPa, Metrologia, 40 (2003) 7002. CrossRef
    13. ISO, Guide to the expression of uncertainty in measurement, ISO Document—ISO/TAG/WG 3, (1995).
    14. NABL, Guidelines for estimation and expression of uncertainty in measurement, NABL Doc. 141(2) (2000).
  • 作者单位:Jasveer Singh (1)
    Nita Dilawar Sharma (1)
    Ashok Kumar (1)
    A. K. Bandyopadhyay (1)

    1. Vacuum and Pressure Standards, CSIR-National Physical Laboratory, Dr. K.S. Krishnan Marg, New Delhi, 110 012, India
  • ISSN:0974-9853
文摘
We report the performance of calibration laboratories in the pneumatic pressure region (0 to 5) MPa through a recently conducted inter-laboratory comparison. Six NABL (National Accreditation Board for Testing and Calibration Laboratories) accredited laboratories participated in the program. The proficiency testing program was organized and piloted by CSIR-National Physical Laboratory. The program started in June 2010 and was completed in October 2011. The artifact used was a high precision pressure dial gauge. The reference values were generated by the pilot laboratory. The deviations for each participating laboratory were estimated against the reference pressure values and the compatibility of the results was calculated using the conventional method. Out of the total measurements made, all but two were found to be in good agreement with the reference values. The normalized error values (E n) of five laboratories out of the total six were found well within?±?1 over the entire pressure range.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700