用户名: 密码: 验证码:
A medical device-grade T1 and ECV phantom for global T1 mapping quality assurance—the T1 Mapping and ECV Standardization in cardiovascular magnetic resonance (T1MES) program
详细信息    查看全文
  • 作者:Gabriella Captur ; Peter Gatehouse…
  • 关键词:T1 mapping ; Standardization ; Phantom
  • 刊名:Journal of Cardiovascular Magnetic Resonance
  • 出版年:2017
  • 出版时间:January 2017
  • 年:2017
  • 卷:18
  • 期:1
  • 全文大小:3,123 KB
  • 刊物主题:Imaging / Radiology; Angiology; Cardiology;
  • 出版者:BioMed Central
  • ISSN:1532-429X
  • 卷排序:18
文摘
BackgroundT1 mapping and extracellular volume (ECV) have the potential to guide patient care and serve as surrogate end-points in clinical trials, but measurements differ between cardiovascular magnetic resonance (CMR) scanners and pulse sequences. To help deliver T1 mapping to global clinical care, we developed a phantom-based quality assurance (QA) system for verification of measurement stability over time at individual sites, with further aims of generalization of results across sites, vendor systems, software versions and imaging sequences. We thus created T1MES: The T1 Mapping and ECV Standardization Program.