用户名: 密码: 验证码:
Antibacterial efficacy and cytotoxicity of low intensity direct current activated silver–titanium implant system prototype
详细信息    查看全文
文摘
Silver-based devices activated by electric current are of interest in biomedicine because of their broad-spectrum antimicrobial activity. This study investigates the in vitro antibacterial efficacy and cytotoxicity of a low intensity direct current (LIDC)-activated silver–titanium implant system prototype designed for localized generation and delivery of silver ions at the implantation site. First, the antibacterial efficacy of the system was assessed against methicillin-resistant Staphylococcus aureus (MRSA) over 48 h at current levels of 3 and 6 µA in Mueller–Hinton broth. The cytotoxicity of the system was then evaluated over 48 h in two phases using an in vitro model with in which the activated electrodes were suspended in growth medium in a cell-seeded tissue culture plate. In phase-1, the system was tested on human osteosarcoma (MG-63) cell line and compared to titanium controls. In phase-2, the cytotoxicity characteristics were validated with normal human diploid osteoblast cells. The LIDC-activated system demonstrated high antimicrobial efficacy against MRSA, but was also toxic to human cells immediately surrounding the electrodes. The statistical analysis showed that the cytotoxicity was a result of the presence of silver, and the electric activation did not make it worse.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号